VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
FDA Z-1018-2026
Medical Device High
Jan 21, 2026 Recall: Dec 19, 2025
Details
ThinPrep¿ CytoLyt¿ Solution. Buffered Cell Wash Solution.
Fungal contamination of affected lot with Parengyodontium album.
FDA Z-1093-2026
Medical Device High
Jan 21, 2026 Recall: Dec 12, 2025
Details
FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA Z-1020-2026
Medical Device High
Jan 21, 2026 Recall: Nov 24, 2025
Details
FDA Z-1084-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
FDA Z-1086-2026
Medical Device High
Jan 21, 2026 Recall: Dec 16, 2025
Details
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
FDA Z-1100-2026
Medical Device High
Jan 21, 2026 Recall: Dec 19, 2025
Details
Allura Xper FD20; Model Number: 722012;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1070-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Allura Xper FD10/10; Model Number: 722005;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1065-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: AN01 CAROTID ANESTH KIT-LF, Medline Kit SKU PHS972096014B
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
FDA Z-1049-2026
Medical Device High
Jan 21, 2026 Recall: Nov 26, 2025
Details
Allura Xper FD20/20; Model Number: 722038;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1077-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
FDA Z-1024-2026
Medical Device High
Jan 21, 2026 Recall: Dec 10, 2025
Details
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
FDA Z-1051-2026
Medical Device High
Jan 21, 2026 Recall: Nov 26, 2025
Details
Allura Xper FD10/10; Model Number: 722011;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1069-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
FDA Z-1040-2026
Medical Device High
Jan 21, 2026 Recall: Dec 4, 2025
Details
FDA Z-1057-2026
Medical Device High
Jan 21, 2026 Recall: Nov 26, 2025
Details
FDA Z-1089-2026
Medical Device High
Jan 21, 2026 Recall: Dec 16, 2025
Details
Allura Xper FD10; Model Number: 722003;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1064-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
FDA Z-1079-2026
Medical Device High
Jan 21, 2026 Recall: Dec 8, 2025
Details
Product Name: Allura Xper FD10C; Model number: 722001;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1062-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Allura Xper FD20; Model Number: 722028;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1075-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
FDA Z-1074-2026
Medical Device High
Jan 21, 2026 Recall: Dec 15, 2025
Details
FDA Z-1055-2026
Medical Device High
Jan 21, 2026 Recall: Nov 26, 2025
Details
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
FDA Z-1039-2026
Medical Device High
Jan 21, 2026 Recall: Oct 29, 2025
Details
A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
FDA Z-1043-2026
Medical Device High
Jan 21, 2026 Recall: Dec 3, 2025
Details
Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
FDA Z-1081-2026
Medical Device High
Jan 21, 2026 Recall: Dec 9, 2025
Details
Showing 2,926 - 2,950 of 3,083 product recalls

What To Enter In The Search Box

A recall usually applies to specific products, batches, model ranges or dates. Use the details printed on the item, package, label, receipt or vehicle paperwork.

Product Name Or Brand

Start broad with the product name, manufacturer, retailer brand or model family. This is usually the fastest way to find a possible notice.

UPC, Lot Code And Dates

For food, drugs, cosmetics and packaged goods, compare the UPC, lot code, product code, best-by date, use-by date and package size.

Model, Serial Or Date Code

For appliances, electronics, toys, furniture and baby products, match the model number, serial range, date code, color, size and retailer listing.

VIN Or NHTSA Campaign

This site can find vehicle recall campaigns. For your exact car, use the official NHTSA VIN lookup to confirm unrepaired recall status.

Official Sources This Checker Organizes

The goal is to make official notices easier to search. The final decision should still come from the linked agency or manufacturer notice.
CPSC

Consumer Product Recalls

Consumer products, toys, furniture, appliances, electronics and baby products. CPSC remedy details can change after a notice is first posted.

FDA

Food And Medical Device Recalls

FDA-regulated foods, drugs, cosmetics and medical devices. FDA notes that not every recall has a press release on the public recalls page.

USDA FSIS

Meat, Poultry And Egg Product Recalls

Meat, poultry and egg product recalls and public health alerts. Check establishment numbers, lot codes, labels and affected states.

NHTSA

Vehicle, Tire And Car Seat Recalls

Vehicle, tire, equipment and child-seat campaigns. VIN or license plate lookup is needed to know whether your specific vehicle still needs repair.

How To Confirm A Recall Match

1. Search The Obvious Name First

Use the brand, product name, manufacturer or retailer. If the result list is long, add a code from the label or package.

2. Open The Closest Result

Read the product description, recall reason, hazard, affected dates, source agency and official recall number before taking action.

3. Match Every Identifier

Compare the UPC, lot, best-by date, establishment number, model, serial range, color, size, state, retailer and product photo.

4. Follow The Official Remedy

The notice may say stop using, do not eat, discard, return for refund, request a repair kit, replace a part or schedule dealer service.

When A Recall Result Looks Like Your Product

Use this checklist before you throw away a product, call the company, request a refund or schedule a repair.

If It Is Food, Formula Or Pet Food

Do not rely on brand alone. Match the package size, UPC, lot code, best-by date, use-by date, distribution states and any allergy or contamination warning.

If It Is A Baby Or Household Product

Look for model number, date code, serial range, color, size, online listing, product photo and the exact hazard such as choking, suffocation, fire or fall risk.

If It Is A Vehicle Or Car Seat

Use campaign results to understand the issue, then verify your VIN, car seat model or tire information with NHTSA or the manufacturer.

If The Details Do Not Match

Your item may not be part of that specific recall. Check spelling, older notices, retailer alerts and the official source before deciding it is safe.

What Is A Product Recall?

A product recall is a safety or compliance notice saying a product may need to be corrected, removed, returned, repaired, replaced, refunded or discarded.

What Should You Do After Finding A Recall?

If all affected details match, follow the official instructions. Keep packaging, take photos, save receipts and contact the company or retailer when a refund or repair is offered.

Why Recall Details Must Match Exactly

A similar brand or product name is not enough. UPCs, lot codes, dates, model numbers, serial ranges and distribution areas determine whether your item is included.

How To Avoid Fake Recall Notices

Be careful with text messages, social posts or refund forms that ask for private information. Verify the notice through an official agency, retailer or manufacturer page.

Product Recall FAQ

Short answers for common product recall searches.

Can I search product recalls by UPC code?

Yes, when the official recall notice includes a UPC or product code. If no result appears, also search the brand, product name, package size, lot code and company name.

Can I search product recalls by lot code?

Yes. Lot codes are especially useful for food, drug, cosmetic and pet food recalls because only certain production batches may be affected.

Does this replace a VIN recall lookup?

No. This tool helps find vehicle recall campaigns and NHTSA numbers. Use the official NHTSA VIN or license plate lookup for vehicle-specific unrepaired recall status.

Are all recalls listed here?

The database combines multiple official U.S. sources, but agencies and companies can publish or update notices at different times. Always check the linked official notice before taking final action.