Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
|
FDA
Z-1695-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 27, 2026
|
Details |
|
Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
|
FDA
Z-1706-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 16, 2026
|
Details |
|
Allura Xper FD20/20 OR Table; System Code: 722039;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
|
FDA
Z-1670-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 3, 2026
|
Details |
|
Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
|
FDA
Z-1648-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
|
Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system.
Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use.
|
FDA
Z-1637-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 18, 2026
|
Details |
|
Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
|
FDA
Z-1680-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 3, 2026
|
Details |
|
eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
|
FDA
Z-1737-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
|
Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and
Due to increased complaints for broken/frayed grip cables for reusable instruments.
|
FDA
Z-1696-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Dec 9, 2025
|
Details |
|
Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
|
FDA
Z-1687-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 18, 2026
|
Details |
|
Burlington Medical, Blockers.
Potential for attenuation degradation over time, decreasing the lifespan.
|
FDA
Z-1764-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 11, 2026
|
Details |
|
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
|
FDA
Z-1649-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
|
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
|
FDA
Z-1647-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
|
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
|
FDA
Z-1683-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 18, 2026
|
Details |
|
MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
|
FDA
Z-1740-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 13, 2026
|
Details |
|
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
|
FDA
Z-1656-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 9, 2026
|
Details |
|
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
|
FDA
Z-1677-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 3, 2026
|
Details |
|
MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
|
FDA
Z-1743-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 13, 2026
|
Details |
|
MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
|
FDA
Z-1741-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 13, 2026
|
Details |
|
Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
|
FDA
Z-1674-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 3, 2026
|
Details |
|
Burlington Medical, Frontal Aprons
Potential for attenuation degradation over time, decreasing the lifespan.
|
FDA
Z-1756-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 11, 2026
|
Details |
|
MiniMed 530G Insulin Pump (MMT-551, MMT-751)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
|
FDA
Z-1751-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 13, 2026
|
Details |
|
Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
Composite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment.
|
FDA
Z-1689-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Jan 28, 2026
|
Details |
|
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O 13. XPS, DYNJ907605
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
|
FDA
Z-1710-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 16, 2026
|
Details |
|
Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation
Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use.
|
FDA
Z-1655-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Jan 29, 2026
|
Details |
|
eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
|
FDA
Z-1735-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
Use brand, product name, UPC, lot code, best-by date, package size, allergen, contamination issue or distribution state.
Cars & equipmentSearch campaigns by make, model, year, component or NHTSA number. Use NHTSA's VIN search for vehicle-specific open recall status.
Family safetyCheck cribs, loungers, strollers, toys, formula, baby food and child car seats by brand, model, date code or hazard.
Fresh updatesBrowse new and recently updated notices across consumer products, foods, vehicles, medical devices and household items.
Start broad with the product name, manufacturer, retailer brand or model family. This is usually the fastest way to find a possible notice.
For food, drugs, cosmetics and packaged goods, compare the UPC, lot code, product code, best-by date, use-by date and package size.
For appliances, electronics, toys, furniture and baby products, match the model number, serial range, date code, color, size and retailer listing.
This site can find vehicle recall campaigns. For your exact car, use the official NHTSA VIN lookup to confirm unrepaired recall status.
Consumer products, toys, furniture, appliances, electronics and baby products. CPSC remedy details can change after a notice is first posted.
FDA-regulated foods, drugs, cosmetics and medical devices. FDA notes that not every recall has a press release on the public recalls page.
Meat, poultry and egg product recalls and public health alerts. Check establishment numbers, lot codes, labels and affected states.
Vehicle, tire, equipment and child-seat campaigns. VIN or license plate lookup is needed to know whether your specific vehicle still needs repair.
Use the brand, product name, manufacturer or retailer. If the result list is long, add a code from the label or package.
Read the product description, recall reason, hazard, affected dates, source agency and official recall number before taking action.
Compare the UPC, lot, best-by date, establishment number, model, serial range, color, size, state, retailer and product photo.
The notice may say stop using, do not eat, discard, return for refund, request a repair kit, replace a part or schedule dealer service.
Do not rely on brand alone. Match the package size, UPC, lot code, best-by date, use-by date, distribution states and any allergy or contamination warning.
Look for model number, date code, serial range, color, size, online listing, product photo and the exact hazard such as choking, suffocation, fire or fall risk.
Use campaign results to understand the issue, then verify your VIN, car seat model or tire information with NHTSA or the manufacturer.
Your item may not be part of that specific recall. Check spelling, older notices, retailer alerts and the official source before deciding it is safe.
A product recall is a safety or compliance notice saying a product may need to be corrected, removed, returned, repaired, replaced, refunded or discarded.
If all affected details match, follow the official instructions. Keep packaging, take photos, save receipts and contact the company or retailer when a refund or repair is offered.
A similar brand or product name is not enough. UPCs, lot codes, dates, model numbers, serial ranges and distribution areas determine whether your item is included.
Be careful with text messages, social posts or refund forms that ask for private information. Verify the notice through an official agency, retailer or manufacturer page.
Yes, when the official recall notice includes a UPC or product code. If no result appears, also search the brand, product name, package size, lot code and company name.
Yes. Lot codes are especially useful for food, drug, cosmetic and pet food recalls because only certain production batches may be affected.
No. This tool helps find vehicle recall campaigns and NHTSA numbers. Use the official NHTSA VIN or license plate lookup for vehicle-specific unrepaired recall status.
The database combines multiple official U.S. sources, but agencies and companies can publish or update notices at different times. Always check the linked official notice before taking final action.