| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
|
FDA
Z-3257-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 3, 2026
|
Details |
|
ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
|
FDA
Z-3250-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 11, 2026
|
Details |
|
Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3237-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
HF2P0ST9 HomeFill Oxygen System HomeFill Cylinder M9 with CGA870 Compatible Post Valve
Labeling; Oxygen Cylinders do not contain required medical gas labeling
|
FDA
Z-3221-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 28, 2026
|
Details |
|
Medline convenience kits containing Healthmark Instrument Channel Brushes. SINUS PACK DYNJ48258D and SINUS PACK DYNJ48258F.
Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of the brush. Potential hazards associated include cut injury/abrasion and infection.
|
FDA
Z-3230-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Intera 1.5T Omni/Stellar (781104), 7. Intera Achieva 1.5T Pulsar (781171);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3240-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-01
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may exhibit an open or incomplete seal on the packaging of the applicator.
|
FDA
Z-3261-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 22, 2026
|
Details |
|
Medline Convenience Kits: 1) MAJOR NEURO VCH, Catalog Number: DYNJ39218I; 2) NEURO PACK-LF, Catalog Number: DYNJ41246F; 3) CRANIOTOMY PACK (CRSJE)227-LF, Catalog Number: DYNJ47487A; 4) MAJOR NEURO PACK, Catalog Number: DYNJ57233J; 5) NEURO, Catalog Number: DYNJ82388A; 6) NEURO PACK, Catalog Number: DYNJ86656; 7) NEURO ACD, Catalog Number: DYNJ904549G
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
|
FDA
Z-3268-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
|
FDA
Z-3139-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 11, 2026
|
Details |
|
SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12
The proximal thread depth of the spinal device is out of design specification.
|
FDA
Z-3226-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
|
FDA
Z-3249-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 7, 2026
|
Details |
|
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0360 with Kit Number: 4-000-0724
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
|
FDA
Z-3220-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Olympus BiCOAG Hemostasis Probe 10Fr Fixed Pin Connector. Model/Catalog Number: CD-B612LA.
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
|
FDA
Z-3245-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Ingenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4. 781271(3.0T CX), 5. 782105(3.0T CX);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3238-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Presource Convenience Kits: 1) KIT, USGPIV , Catalog Number: PVMHIVMHD; 2) KIT, USGPIV , Catalog Number: PVMHIVM14; 3) KIT, USGPIV , Catalog Number: PVMHIVM15
The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.
|
FDA
Z-3255-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;
The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.
|
FDA
Z-3224-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 7, 2026
|
Details |
|
Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3235-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
REF:HF2PC9 HomeFill Oxygen System, HomeFill Cylinder M9 with Integrated Conserver
Labeling; Oxygen Cylinders do not contain required medical gas labeling
|
FDA
Z-3222-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 28, 2026
|
Details |
|
Brand Name: Pearle Vision 1 day Product Name: (somofilcon A) Contact Lens Model/Catalog Number: Pearle Vision 1 day
A small number of lenses within the affected lot are of the incorrect power.
|
FDA
Z-3248-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 12, 2026
|
Details |
|
Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5; M001314021 IMAGER II/5/ST/65/035 BX 5; M001314031 IMAGER II/5/ST/100/035 BX 5; M001314041 IMAGER II/5/BARNH/100/038 BX5; M001314051 IMAGER II/5/BERN/100/038 BX5; M001314061 IMAGER II/5/BERN/100/035 BX 5; M001314081 IMAGER II/5/BERN/40/038 BX5; M001314101 IMAGER II/5/BERN/65/038 BX5; M001314141 IMAGER II/5/H1/100/038 BX5; M001314151 IMAGER II/5/H1/100/035 BX 5; M001314301 IMAGER II/5/SIM1/65/038 BX5; M001314321 IMAGER II/5/SIM1/100/038 BX5; M001314341 IMAGER II/5/SIM2/100/038 BX5; M0013
Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
|
FDA
Z-3214-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 24, 2026
|
Details |
|
Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3) KIT, PRE PROCEDURE , Catalog Number: IVSJV8302E; 4) KIT, IV SUPPLIES , Catalog Number: IVS3223531; 5) KIT, IV SUPPLIES , Catalog Number: IVS3223532; 6) KIT, IV SUPPLIES , Catalog Number: IVS3223533; 7) KIT, IV SUPPLIES , Catalog Number: IVS3223534; 8) KIT, IV SUPPLIES , Catalog Number: IVS3223535; 9) KIT, IV SUPPLIES , Catalog Number: IVS322358M; 10) KIT, IV SUPPLIES , Catalog Number: IVS322358N; 11) KIT, IV SUPPLIES , Catalog Number: IVS322358O; 12) KIT, PATIENT
The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.
|
FDA
Z-3254-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
|
FDA
Z-3246-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may exhibit an open or incomplete seal on the packaging of the applicator.
|
FDA
Z-3260-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 22, 2026
|
Details |
|
PRO+ 40" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned, Model/Catalog Number: P7924A14
There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame.
|
FDA
Z-3264-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 18, 2026
|
Details |
|
Achieva XR; Model Numbers (REF): 1. 781153, 2. 781253;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3234-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.