| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
REF: 81-030415 Myocardial Probe Non-Pyrogenic , 400 Series
Myocardial temperature probes contain an incorrect needle. Specifically, an 18mm needle was included instead of a 15mm needle.
|
FDA
Z-3256-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 20, 2026
|
Details |
|
Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit Model Number: KRIV26-02.
Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.
|
FDA
Z-3194-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 22, 2026
|
Details |
|
New York Embroidery Studio Trinity Non-Surgical Isolation Gown, AAMI PB70 Level 2, X-Large. Model Number: NYES100-2. Packaged in inner packs of 10 and 100 gowns per shipping case.
Gowns do not consistently meet the minimum liquid-barrier performance required.
|
FDA
Z-3259-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 31, 2026
|
Details |
|
REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve
Labeling; Oxygen Cylinders do not contain required medical gas labeling
|
FDA
Z-3223-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 28, 2026
|
Details |
|
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye
Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
|
FDA
Z-3215-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 6, 2026
|
Details |
|
Medline Convenience Kits: 1) ANTERIOR SPINE PACK, Catalog Number: DYNJT10197; 2) CRANIOTOMY #IMF 56646-LF, Catalog Number: DYNJVB91001; 3) CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348O; 4) SLUH CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348P; 5) LAMINECTOMY PACK-LF, Catalog Number: DYNJ04342M; 6) LAMINECTOMY PACK-LF, Catalog Number: DYNJ43218G; 7) CRANIOTOMY PACK, Catalog Number: DYNJ46024F; 8) LAMINECTOMY PACK-LF, Catalog Number: DYNJ46070C; 9) CRANI PACK, Catalog Number: DYNJ46672B; 10) SPINE TRAY II, Catalog Number: DYNJ52961I; 11) SPINE PACK, Catalog Number: DYNJ53518C; 12) SPINE PACK,
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
|
FDA
Z-3269-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
|
FDA
Z-3216-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Ingenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3. 781396(1.5T), 4. 782101(1.5T), 5. 782115(1.5T), 6. 782140(1.5T), 7. 781261(1.5T CX), 8. 781262(1.5T CX), 9. 781347(1.5T S);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3236-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
|
FDA
Z-3244-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Cardioroot Woven Collagen Coated Vascular Graft; Model Numbers: 1. HEWROOT0024, 2. HEWROOT0026, 3. HEWROOT0028, 4. HEWROOT0030, 5. HEWROOT0032, 6. HEWROOT0034;
Product failed to meet the specified acceptance criteria for Integral Water Permeability (IWP).
|
FDA
Z-3258-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 12, 2026
|
Details |
|
MR 5300 Model Number (REF): 1. 782110, 2. 782152;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3241-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3242-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
|
FDA
Z-3231-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 24, 2026
|
Details |
|
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0320 with Kit Number: 4-000-0622
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
|
FDA
Z-3217-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
|
FDA
Z-3253-2026
|
Medical Device | High |
Sep 30, 2026
Recall: May 6, 2026
|
Details |
|
Ingenia Elition S and Ingenia Elition X; Model Numbers (REF): 1. 781357(S), 2. 782106(S), 3. 782137(S), 4. 781358(X), 5. 782107(X), 6. 782119(X), 7. 782136(X), 8. 782158(X);
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3239-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1;
Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.
|
FDA
Z-3225-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Jul 29, 2026
|
Details |
|
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-0350 with Kit Number: 4-000-0722
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
|
FDA
Z-3219-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
|
FDA
Z-3140-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 11, 2026
|
Details |
|
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
|
FDA
Z-3251-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Feb 19, 2019
|
Details |
|
Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
|
FDA
Z-3233-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-03300 with Kit Number: 4-000-0624
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
|
FDA
Z-3218-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 21, 2026
|
Details |
|
Medline convenience kits containing Healthmark Instrument Channel Brushes. SINUS PACK DYNJ48258D and Sinus Pack DYNJ48258F.
Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of the brush. Potential hazards associated include cut injury/abrasion and infection.
|
FDA
Z-3229-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 10, 2026
|
Details |
|
Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA.
Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
|
FDA
Z-3247-2026
|
Medical Device | High |
Sep 30, 2026
Recall: Aug 19, 2026
|
Details |
|
Convenience kits CATARACT PACK DYNJ59881A
Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.
|
FDA
Z-3210-2026
|
Medical Device | High |
Sep 23, 2026
Recall: Jul 31, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.