Burlington Medical, Frontal Aprons
Potential for attenuation degradation over time, decreasing the lifespan.
FDA Z-1756-2026
Medical Device High
Apr 8, 2026 Recall: Feb 11, 2026
Details
MiniMed 530G Insulin Pump (MMT-551, MMT-751)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
FDA Z-1751-2026
Medical Device High
Apr 8, 2026 Recall: Feb 13, 2026
Details
Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011.
Composite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment.
FDA Z-1689-2026
Medical Device High
Apr 8, 2026 Recall: Jan 28, 2026
Details
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O 13. XPS, DYNJ907605
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
FDA Z-1710-2026
Medical Device High
Apr 8, 2026 Recall: Feb 16, 2026
Details
FDA Z-1655-2026
Medical Device High
Apr 8, 2026 Recall: Jan 29, 2026
Details
eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
FDA Z-1735-2026
Medical Device High
Apr 8, 2026 Recall: Mar 2, 2026
Details
ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
FDA Z-1665-2026
Medical Device High
Apr 8, 2026 Recall: Mar 3, 2026
Details
ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
FDA Z-1690-2026
Medical Device High
Apr 8, 2026 Recall: Feb 13, 2026
Details
FDA Z-1699-2026
Medical Device High
Apr 8, 2026 Recall: Dec 9, 2025
Details
Burlington Medical, Protective Sleeves.
Potential for attenuation degradation over time, decreasing the lifespan.
FDA Z-1759-2026
Medical Device High
Apr 8, 2026 Recall: Feb 11, 2026
Details
FDA Z-1686-2026
Medical Device High
Apr 8, 2026 Recall: Feb 18, 2026
Details
Allura Xper FD20/10; System Code: 722029;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
FDA Z-1668-2026
Medical Device High
Apr 8, 2026 Recall: Mar 3, 2026
Details
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc
Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break.
FDA Z-1682-2026
Medical Device High
Apr 8, 2026 Recall: Feb 18, 2026
Details
Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
FDA Z-1709-2026
Medical Device High
Apr 8, 2026 Recall: Feb 16, 2026
Details
FDA Z-1589-2026
Medical Device High
Apr 8, 2026 Recall: Feb 18, 2026
Details
Burlington Medical, Kilt Blockers.
Potential for attenuation degradation over time, decreasing the lifespan.
FDA Z-1765-2026
Medical Device High
Apr 8, 2026 Recall: Feb 11, 2026
Details
Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT
Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.
FDA Z-1691-2026
Medical Device High
Apr 8, 2026 Recall: Feb 13, 2026
Details
Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
FDA Z-1750-2026
Medical Device High
Apr 8, 2026 Recall: Feb 13, 2026
Details
FDA Z-1681-2026
Medical Device High
Apr 8, 2026 Recall: Feb 18, 2026
Details
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
FDA Z-1657-2026
Medical Device High
Apr 8, 2026 Recall: Feb 23, 2026
Details
ALLURA Xper FD10F; System Code: 722002;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
FDA Z-1660-2026
Medical Device High
Apr 8, 2026 Recall: Mar 3, 2026
Details
Burlington Medical, Demi Half Aprons.
Potential for attenuation degradation over time, decreasing the lifespan.
FDA Z-1763-2026
Medical Device High
Apr 8, 2026 Recall: Feb 11, 2026
Details
Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
FDA Z-1694-2026
Medical Device High
Apr 8, 2026 Recall: Feb 27, 2026
Details
ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
FDA Z-1666-2026
Medical Device High
Apr 8, 2026 Recall: Mar 3, 2026
Details
FDA Z-1591-2026
Medical Device High
Apr 8, 2026 Recall: Feb 16, 2026
Details
Showing 976 - 1,000 of 1,678 product recalls

Medical Device Recalls Today

Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.

Search The Device Label

Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.

Recall Class And Risk Level

Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.

What To Check On A Device Label

If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.

How To Check Medical Device Recalls: FDA Device Safety Alerts And Recall Notices

Match The Exact Medical device

Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.

Read The Recall Hazard

Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.

Check Where It Was Sold

Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.

Follow The Official Remedy

Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.

Medical Device Recalls: FDA Device Safety Alerts And Recall Notices FAQ

Short answers for checking whether a specific item is included in this recall category.

How do I know if this medical device is recalled?

Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.

What should I do after finding a matching recall?

Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.

Why can two similar products have different recall status?

Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.

Is this page the official recall notice?

No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.