| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
|
FDA
Z-1734-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Mar 2, 2026
|
Details |
|
MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
|
FDA
Z-1745-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 13, 2026
|
Details |
|
Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Number: 420183 Software Version: NA Product Description: EndoWrist Instruments or da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
|
FDA
Z-1701-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Dec 9, 2025
|
Details |
|
SIGNA Premier systems
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
|
FDA
Z-1767-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 27, 2026
|
Details |
|
THORACIC ROBOTS, DYNJ908777B
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
|
FDA
Z-1707-2026
|
Medical Device | High |
Apr 8, 2026
Recall: Feb 16, 2026
|
Details |
|
Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
|
FDA
Z-1614-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Mar 2, 2026
|
Details |
|
Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
|
FDA
Z-1650-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 10, 2026
|
Details |
|
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
|
FDA
Z-1600-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jan 23, 2026
|
Details |
|
Philips L12-5 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
|
FDA
Z-1620-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Sep 5, 2025
|
Details |
|
Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
|
FDA
Z-1641-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 25, 2026
|
Details |
|
MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Due to the likely presence of contamination in well(s).
|
FDA
Z-1609-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jul 24, 2025
|
Details |
|
Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
|
FDA
Z-1592-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jan 23, 2026
|
Details |
|
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Probes may rupture/burst during activation
|
FDA
Z-1567-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 12, 2026
|
Details |
|
Philips C9-4 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
|
FDA
Z-1619-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Sep 5, 2025
|
Details |
|
Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.
Product labeling includes a fetal indication for use that is not cleared under its 510(k).
|
FDA
Z-1631-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Mar 4, 2026
|
Details |
|
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
|
FDA
Z-1644-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 25, 2026
|
Details |
|
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
|
FDA
Z-1639-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 25, 2026
|
Details |
|
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
|
FDA
Z-1596-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jan 23, 2026
|
Details |
|
Philips X7-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
|
FDA
Z-1628-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Sep 5, 2025
|
Details |
|
Philips 3D6-2 Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
|
FDA
Z-1618-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Sep 5, 2025
|
Details |
|
FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
|
FDA
Z-1595-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jan 23, 2026
|
Details |
|
Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
|
FDA
Z-1633-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Mar 17, 2026
|
Details |
|
Embrace Drill Tower, Standard/Lateral (25mm); Item Number: 645-081/63;
The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct.
|
FDA
Z-1652-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Feb 11, 2026
|
Details |
|
I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
|
FDA
Z-1604-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Jan 23, 2026
|
Details |
|
Philips OMNI III TEE Ultrasound Transducer
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
|
FDA
Z-1624-2026
|
Medical Device | High |
Apr 1, 2026
Recall: Sep 5, 2025
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.