| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
Artelon FlexBand Dynamic Matrix Ref: 31057
Augmentation devices failed bacterial endotoxin testing.
|
FDA
Z-1554-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Feb 6, 2026
|
Details |
|
Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
|
FDA
Z-1546-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Feb 6, 2026
|
Details |
|
CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
|
FDA
Z-1553-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 30, 2026
|
Details |
|
GEM Premier 5000 PAK, Part No. 00055360004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1524-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
Penner Pacific Bathing Spa, Model Numbers 360020-1P
The device does not bear a unique device identifier.
|
FDA
Z-1502-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Aug 13, 2025
|
Details |
|
Endo-Model Replacement Plateau; Item Number: 15-0027/11;
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
|
FDA
Z-1521-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 12, 2026
|
Details |
|
Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
|
FDA
Z-1478-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Nov 14, 2025
|
Details |
|
Brand Name: HiResolution Bionic Ear System Product Name: Sky CI M90 Sound Processor Model/Catalog Number: CI-5295-150
Behind-the-ear sound processer packaging label is different then included product.
|
FDA
Z-1493-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Nov 27, 2025
|
Details |
|
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various si
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
|
FDA
Z-1494-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 8, 2026
|
Details |
|
Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE, 636007PE, 631512PE, 636012PE, 636014PE, 632211P50E, 633016PE, 631407PE, 631809PE, 632010PE, 631608P50E, 632520PE, 632412PE, 631510PE, 635022PE, 631224PE, 632412P50E, 631507PE, 633014PE, 632010P50E, 635020PE, 631809P50E. Gelweave, REF: 736038E, 736036E, 736034E, 736032ST/10DE, 736032E, 736030ST/10DE
Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.
|
FDA
Z-1492-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Dec 18, 2025
|
Details |
|
Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable
The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
|
FDA
Z-1491-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Feb 4, 2026
|
Details |
|
Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
|
FDA
Z-1477-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Nov 14, 2025
|
Details |
|
Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
|
FDA
Z-1481-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Feb 2, 2026
|
Details |
|
Brand Name: B. Braun Product Name: 21GA WINGED INFUSION, SINGLE PK Model/Catalog Number: 7A3842 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
|
FDA
Z-1482-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Feb 2, 2026
|
Details |
|
CS100 IABP. Software Version CS100 IABP Q.01.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
|
FDA
Z-1486-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 23, 2026
|
Details |
|
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
|
FDA
Z-1484-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 20, 2026
|
Details |
|
CS300 IABP. Software Version CS300 IABP C.01.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
|
FDA
Z-1487-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 23, 2026
|
Details |
|
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological
Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,
|
FDA
Z-1475-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 16, 2026
|
Details |
|
cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
|
FDA
Z-1476-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 20, 2026
|
Details |
|
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.
|
FDA
Z-1490-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Mar 14, 2025
|
Details |
|
Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A
There is the potential for the length of the trocar shaft to be too long.
|
FDA
Z-1485-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 15, 2026
|
Details |
|
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
|
FDA
Z-1480-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 21, 2026
|
Details |
|
Impella RP. Product Code: 0046-0011.
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
|
FDA
Z-1471-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Jan 27, 2026
|
Details |
|
Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product Description: A device designed for subcutaneous implantation and intended to be used temporarily to expand an implanted tissue expander by percutaneous injection of air or fluids. Component: N/A
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break.
|
FDA
Z-1483-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Feb 2, 2026
|
Details |
|
Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
|
FDA
Z-1479-2026
|
Medical Device | High |
Mar 11, 2026
Recall: Nov 14, 2025
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.