| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD15
Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.
|
FDA
Z-2294-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 20, 2026
|
Details |
|
BD Pyxis Anesthesia Station ES REF: 327 Medication cabinet
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
|
FDA
Z-2299-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 30, 2026
|
Details |
|
The Large Volume Pump (LVP) of the Ivenix Infusion System (IIS). Model Number: LVP-0004.
Potential for the Cassette Loading Lever to break.
|
FDA
Z-2274-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Feb 4, 2026
|
Details |
|
BD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405609.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
|
FDA
Z-2248-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
|
FDA
Z-2205-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Mar 9, 2026
|
Details |
|
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405658.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
|
FDA
Z-2254-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in., BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405741.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
|
FDA
Z-2253-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); 25GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333862); GOVP24BK 24G PENCAN TRAY W/ BUP & DEXTROS (332607); P24BK 24G PENCAN TRAY W/BUP & DEXTROSE (333868); P25BKAY PENCAN SPINAL TRAY (333865); P27BK PENCAN SPINAL W/BUPIVAC (333871); P25BK SPINAL TRAY (333851)
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
|
FDA
Z-2276-2026
|
Medical Device | High |
Jun 10, 2026
Recall: May 6, 2026
|
Details |
|
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
|
FDA
Z-2275-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
BD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
|
FDA
Z-2249-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 405723, 406704.
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing.
|
FDA
Z-2255-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 27, 2026
|
Details |
|
TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis
Due to incorrect functional length on device labeling.
|
FDA
Z-2308-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 3, 2026
|
Details |
|
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2291-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
|
FDA
Z-2206-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Mar 9, 2026
|
Details |
|
Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2287-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: BPSK, BLOCKJOCKS TRAY Model/Catalog Number: 570246 Software Version: N/A Product Description: BPSK, BLOCKJOCKS TRAY Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2264-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
BD Pyxis MedStation ES REF: 323 Medication cabinet
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
|
FDA
Z-2298-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 30, 2026
|
Details |
|
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2290-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices are intended for use in providing tracheal access for airway management. Shiley Tracheostomy Tubes are double cannula tracheostomy tubes with reusable inner cannulae (LPC, FEN, CFS, CFN, LGT) and twist-lock connectors. These Shiley Tracheostomy Tubes have a radiopaque, biocompatible outer cannula constructed of polyvinyl chloride.
Due to customer complaint regarding incorrect display box labeling.
|
FDA
Z-2261-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 15, 2026
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: NES1725KFXL SPINAL EPID CLOSED TIP CATH Model/Catalog Number: 530201 Software Version: N/A Product Description: NES1725KFXL SPINAL EPID CLOSED TIP CATH Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2265-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
FDA
Z-2283-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 30, 2026
|
Details | |
|
BD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
|
FDA
Z-2300-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 30, 2026
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK SHANDS JACKSONVILLE 20055717 Model/Catalog Number: 555884 Software Version: N/A Product Description: CESK SHANDS JACKSONVILLE 20055717 Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2269-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
|
FDA
Z-2306-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 21, 2026
|
Details |
|
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2288-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.