| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor
Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
|
FDA
Z-1795-2026
|
Medical Device | High |
Apr 15, 2026
Recall: Mar 12, 2026
|
Details |
|
COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
|
FDA
Z-1778-2026
|
Medical Device | High |
Apr 15, 2026
Recall: Mar 5, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055407504.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1528-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055445010 & 00055445011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1543-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055415004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1532-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055360010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1526-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055430011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1540-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00,
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
|
FDA
Z-1545-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Feb 6, 2026
|
Details |
|
GEM Premier 5000; Part No: 00055415011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1536-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Pat
Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
|
FDA
Z-1557-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 30, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055430004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1537-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055415010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1535-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055445004.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1541-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i8
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
|
FDA
Z-1548-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Feb 6, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055360008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1525-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055407510.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1530-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055445008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1542-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055407508.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1529-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
DxC 700 AU, REF: B86444, B86446
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
|
FDA
Z-1552-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 30, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055360011.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1527-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000 PAK; Part No. 00055407511.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1531-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055415008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1534-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055430008.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1538-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Model No. 00055430010.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1539-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
|
GEM Premier 5000; Part No. 00055415005.
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
|
FDA
Z-1533-2026
|
Medical Device | High |
Mar 18, 2026
Recall: Jan 20, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.