| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
|
FDA
Z-2362-2026
|
Medical Device | High |
Jun 17, 2026
Recall: Apr 16, 2026
|
Details |
|
SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
|
FDA
Z-2353-2026
|
Medical Device | High |
Jun 17, 2026
Recall: May 4, 2026
|
Details |
|
Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2286-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: SESK UPMC DOUBLE CATHETER TRAY Model/Catalog Number: 530134 Software Version: N/A Product Description: SESK UPMC DOUBLE CATHETER TRAY Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2270-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK RIVERSIDE HEALTH SYSTEM Model/Catalog Number: 552126 Software Version: N/A Product Description: CESK RIVERSIDE HEALTH SYSTEM Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2268-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
|
FDA
Z-2223-2026
|
Medical Device | High |
Jun 10, 2026
Recall: May 6, 2026
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: LYNDON B JOHNSON HOSPITAL SPINAL EPID Model/Catalog Number: 530184 Software Version: N/A Product Description: LYNDON B JOHNSON HOSPITAL SPINAL EPID Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2267-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2289-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2291-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2287-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: BPSK, BLOCKJOCKS TRAY Model/Catalog Number: 570246 Software Version: N/A Product Description: BPSK, BLOCKJOCKS TRAY Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2264-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2290-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: NES1725KFXL SPINAL EPID CLOSED TIP CATH Model/Catalog Number: 530201 Software Version: N/A Product Description: NES1725KFXL SPINAL EPID CLOSED TIP CATH Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2265-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
FDA
Z-2283-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 30, 2026
|
Details | |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK SHANDS JACKSONVILLE 20055717 Model/Catalog Number: 555884 Software Version: N/A Product Description: CESK SHANDS JACKSONVILLE 20055717 Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2269-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
|
FDA
Z-2306-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 21, 2026
|
Details |
|
Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2288-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use wi
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
|
FDA
Z-2310-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Apr 24, 2026
|
Details |
|
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2284-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC Product Name: CESK NORTHSIDE ANESTHESIA TRAY Model/Catalog Number: 555076 Software Version: N/A Product Description: CESK NORTHSIDE ANESTHESIA TRAY Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2263-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
|
FDA
Z-2285-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Dec 7, 2023
|
Details |
|
Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK HARRIS METHODIST Model/Catalog Number: 555498 Software Version: N/A Product Description: CESK HARRIS METHODIST Component: No
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL¿ Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
|
FDA
Z-2266-2026
|
Medical Device | High |
Jun 10, 2026
Recall: Jan 29, 2026
|
Details |
|
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2243-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2247-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Number: 11-9099 Software Version: No Product Description: 11-9099 - ICEMAN, W/COLDPAD, NS, RH The IceMan CLASSIC3¿ cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan¿ helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
|
FDA
Z-2137-2026
|
Medical Device | High |
May 20, 2026
Recall: Mar 19, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.