FDA device recall and safety alerts

Medical Device Recalls: FDA Device Safety Alerts And Recall Notices

Search FDA medical device recalls and early alerts for equipment, software, implants, monitors, diagnostic tools and patient-use devices. Match device name, company, recall class, product code, UDI, model, lot and serial details.

Search Medical Device Recalls By Device, Company Or Recall Class

Use this medical device recall search to check FDA device notices by device name, company, product code, lot number, recall class, risk level or correction/removal details.

Medical Device Recall Alerts List

Showing 51 - 75 of 2,579 matching product recalls.
Product Recall Search
Product Recall Source Type Risk Date Action
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0261-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/155; Material Number: M001195480
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0246-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
MEDLINE medical convenience kits labeled as: C-SECTION PACK, Medline Product Number/SKU DYNJ48455A
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
FDA Z-0273-2027
Medical Device High
Oct 28, 2026 Recall: Sep 9, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/130; Material Number: M001195640
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0251-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
MEDLINE medical convenience kits labeled as: TUBAL LIGATION RFD, Medline Product Number/SKU DYNJ906291G
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
FDA Z-0279-2027
Medical Device High
Oct 28, 2026 Recall: Sep 9, 2026
Details
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0242-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.
A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
FDA Z-0282-2027
Medical Device High
Oct 28, 2026 Recall: Sep 8, 2026
Details
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
FDA Z-0202-2027
Medical Device High
Oct 28, 2026 Recall: Sep 21, 2026
Details
FDA Z-0191-2027
Medical Device High
Oct 28, 2026 Recall: Sep 10, 2026
Details
MEDLINE DRAPE C-SEC FEN CLEAR SCRN WIRE, Medline Product Number/SKU DYNJP6106
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
FDA Z-0270-2027
Medical Device High
Oct 28, 2026 Recall: Sep 9, 2026
Details
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/155; Material Number: M001195510
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0248-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0245-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
FDA Z-0205-2027
Medical Device High
Oct 28, 2026 Recall: Aug 14, 2026
Details
VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
FDA Z-0130-2027
Medical Device High
Oct 28, 2026 Recall: Sep 17, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/2RO/130; Material Number: M001195660
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0253-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/2RO/130; Material Number: M001195670
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0254-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Philips Lumify Diagnostic Ultrasound System
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
FDA Z-0152-2027
Medical Device High
Oct 28, 2026 Recall: Sep 24, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/130; Material Number: M001195610
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0249-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0250-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
FDA Z-0125-2027
Medical Device High
Oct 28, 2026 Recall: Sep 17, 2026
Details
Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly
Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
FDA Z-0195-2027
Medical Device High
Oct 28, 2026 Recall: Aug 31, 2026
Details
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0239-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
FDA Z-0123-2027
Medical Device High
Oct 28, 2026 Recall: Sep 17, 2026
Details
MEDLINE medical convenience kits labeled as: C-SECTION TRAY-LF, Medline Kit Number/SKU OBN471K
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
FDA Z-0280-2027
Medical Device High
Oct 28, 2026 Recall: Sep 9, 2026
Details
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/155; Material Number: M001195720
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
FDA Z-0256-2027
Medical Device High
Oct 28, 2026 Recall: Sep 16, 2026
Details
Showing 51 - 75 of 2,579 product recalls

Medical Device Recalls Today

Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.

Search The Device Label

Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.

Recall Class And Risk Level

Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.

What To Check On A Device Label

If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.

How To Check Medical Device Recalls: FDA Device Safety Alerts And Recall Notices

Match The Exact Medical device

Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.

Read The Recall Hazard

Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.

Check Where It Was Sold

Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.

Follow The Official Remedy

Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.

Medical Device Recalls: FDA Device Safety Alerts And Recall Notices FAQ

Short answers for checking whether a specific item is included in this recall category.

How do I know if this medical device is recalled?

Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.

What should I do after finding a matching recall?

Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.

Why can two similar products have different recall status?

Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.

Is this page the official recall notice?

No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.