FreeStyle Libre 3 Plus Sensor REF 72080-01 UDI-DI code: 00357599819002 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA Z-1023-2026
Medical Device High
Jan 21, 2026 Recall: Nov 24, 2025
Details
FreeStyle Libre 3 Plus Sensor REF 78769-01 UDI-DI code: 00357599843014 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA Z-1022-2026
Medical Device High
Jan 21, 2026 Recall: Nov 24, 2025
Details
FDA Z-1095-2026
Medical Device High
Jan 21, 2026 Recall: Dec 11, 2025
Details
FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA Z-1021-2026
Medical Device High
Jan 21, 2026 Recall: Nov 24, 2025
Details
FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
FDA Z-1020-2026
Medical Device High
Jan 21, 2026 Recall: Nov 24, 2025
Details
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
FDA Z-1079-2026
Medical Device High
Jan 21, 2026 Recall: Dec 8, 2025
Details
FDA Z-1037-2026
Medical Device High
Jan 14, 2026 Recall: Dec 12, 2025
Details
Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
FDA Z-1033-2026
Medical Device High
Jan 14, 2026 Recall: Dec 12, 2025
Details
Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
FDA Z-1027-2026
Medical Device High
Jan 14, 2026 Recall: Nov 26, 2025
Details
FDA Z-1029-2026
Medical Device High
Jan 14, 2026 Recall: Nov 17, 2025
Details
FDA Z-1036-2026
Medical Device High
Jan 14, 2026 Recall: Dec 12, 2025
Details
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
FDA Z-1008-2026
Medical Device High
Jan 7, 2026 Recall: Dec 5, 2025
Details
Showing 76 - 87 of 87 product recalls

Medical Device Recalls Today

Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.

Search The Device Label

Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.

Recall Class And Risk Level

Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.

What To Check On A Device Label

If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.

How To Check Medical Device Recalls: FDA Device Safety Alerts And Recall Notices

Match The Exact Medical device

Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.

Read The Recall Hazard

Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.

Check Where It Was Sold

Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.

Follow The Official Remedy

Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.

Medical Device Recalls: FDA Device Safety Alerts And Recall Notices FAQ

Short answers for checking whether a specific item is included in this recall category.

How do I know if this medical device is recalled?

Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.

What should I do after finding a matching recall?

Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.

Why can two similar products have different recall status?

Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.

Is this page the official recall notice?

No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.