Medical Device Recalls Today
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Search FDA medical device recalls and early alerts for equipment, software, implants, monitors, diagnostic tools and patient-use devices. Match device name, company, recall class, product code, UDI, model, lot and serial details.
Use this medical device recall search to check FDA device notices by device name, company, product code, lot number, recall class, risk level or correction/removal details.
| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
Allura Xper FD10/10; Model Number: 722005;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1065-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: AN01 CAROTID ANESTH KIT-LF, Medline Kit SKU PHS972096014B
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
|
FDA
Z-1049-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 26, 2025
|
Details |
|
Allura Xper FD20/20; Model Number: 722038;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1077-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
|
FDA
Z-1024-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 10, 2025
|
Details |
|
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DELUXE PRE OP KIT, Medline SKU DYKS2034A; 2) DELUXE PRE OP KIT, Medline SKU DYKS2006B.
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
|
FDA
Z-1051-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 26, 2025
|
Details |
|
Allura Xper FD10/10; Model Number: 722011;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1069-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System.
AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
|
FDA
Z-1040-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 4, 2025
|
Details |
|
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
|
FDA
Z-1057-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 26, 2025
|
Details |
|
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
|
FDA
Z-1089-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 16, 2025
|
Details |
|
Allura Xper FD10; Model Number: 722003;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1064-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
|
FDA
Z-1079-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 8, 2025
|
Details |
|
Product Name: Allura Xper FD10C; Model number: 722001;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1062-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Allura Xper FD20; Model Number: 722028;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1075-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1074-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DRAWER 2 IV SUPPLIES, Medline kit SKU ACC010271B; 2) DRAWER 2 IV FLUIDS & TUBING, Medline kit SKU ACC010416B; 3) OXYGENATION & TUBING TRAY, Medline kit SKU ACC010525C; 4) TOP DRAWER #6, Medline kit SKU ACC010688; 5) RRMC CODE CART DRAWER 3, Medline kit SKU ACC010770; 6) SURGERY BAG, Medline kit SKU DYKM1279C; 7) ADMISSION KIT, Medline kit SKU DYKS1179; 8) PRE OP KIT, Medline kit SKU DYKS1192C; 9) MEDLINE PEDIATRIC PATIENT PACK, Medline kit SKU DYKS1199B; 10) MEDLINE ADULT PATIENT PACKAGE, Medli
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
|
FDA
Z-1055-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 26, 2025
|
Details |
|
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
|
FDA
Z-1039-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Oct 29, 2025
|
Details |
|
A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
|
FDA
Z-1043-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 3, 2025
|
Details |
|
Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
|
FDA
Z-1081-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 9, 2025
|
Details |
|
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL PACK, Medline kit SKU DYNJ54488A
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
|
FDA
Z-1054-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 26, 2025
|
Details |
|
Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
|
FDA
Z-1063-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
BD Ready-to-Use Hydrophilic Catheter 14 Fr (4.7 mm) Straight Tip, Coated, Female REF: RTU14F
Labeling error, Incorrect expiration date
|
FDA
Z-1061-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Nov 29, 2025
|
Details |
|
da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,Box)
Access Port System tray may develop cracks potentially resulting in a sterility breach.
|
FDA
Z-1096-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 15, 2025
|
Details |
|
Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
|
FDA
Z-1090-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 16, 2025
|
Details |
|
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
|
FDA
Z-1017-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 19, 2025
|
Details |
|
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
|
FDA
Z-1078-2026
|
Medical Device | High |
Jan 21, 2026
Recall: Dec 10, 2025
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.