Medical Device Recalls Today
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Search FDA medical device recalls and early alerts for equipment, software, implants, monitors, diagnostic tools and patient-use devices. Match device name, company, recall class, product code, UDI, model, lot and serial details.
Use this medical device recall search to check FDA device notices by device name, company, product code, lot number, recall class, risk level or correction/removal details.
| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
|
FDA
Z-3130-2026
|
Medical Device | High |
Sep 23, 2026
Recall: Jul 29, 2026
|
Details |
|
StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
|
FDA
Z-3131-2026
|
Medical Device | High |
Sep 23, 2026
Recall: Jul 29, 2026
|
Details |
|
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
|
FDA
Z-3192-2026
|
Medical Device | High |
Sep 23, 2026
Recall: Jul 30, 2026
|
Details |
|
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
|
FDA
Z-3110-2026
|
Medical Device | High |
Sep 16, 2026
Recall: Aug 6, 2026
|
Details |
|
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7)
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
|
FDA
Z-3082-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 6, 2026
|
Details |
|
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS ANZ, Model Number 7248, All Versions (Current 7.7)
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
|
FDA
Z-3083-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 6, 2026
|
Details |
|
Philips Brilliance. Models: Brilliance iCT, Brilliance iCT Upgrades, Brilliance iCT SP. Software 4.1.10.x Product Codes: 728306, 728305, 728311.
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
|
FDA
Z-3040-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 7, 2026
|
Details |
|
Philips IQon Spectral CT Product Codes: IQon Spectral CT 728332, IQon Spectral CT 728331 Software 4.7.7.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
|
FDA
Z-3041-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 7, 2026
|
Details |
|
Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL
A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
|
FDA
Z-3066-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 11, 2026
|
Details |
|
Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS
A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
|
FDA
Z-3065-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 11, 2026
|
Details |
|
Velys Enabling Tech Robotic-Assist Station REF: 451611001
Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
|
FDA
Z-3088-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Jul 23, 2026
|
Details |
|
Philips Ingenuity CT Product Codes: Ingenuity CT Upgrades 728327, Ingenuity CT China 728324, Ingenuity Core 128 Elite Brazil 728325. Software 4.4.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
|
FDA
Z-3042-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 7, 2026
|
Details |
|
Philips Brilliance CT 64: Brilliance CT 64 Channel Product Code: 728231, Brilliance CT 64 Upgrades Product Code: 728232. Software 4.1.10.x
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
|
FDA
Z-3043-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 7, 2026
|
Details |
|
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model Number 7260, All Versions (Current 7.8)
There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
|
FDA
Z-3081-2026
|
Medical Device | High |
Sep 9, 2026
Recall: Aug 6, 2026
|
Details |
|
PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
|
FDA
Z-3036-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
Philips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.
Potential of failing to trigger clinical scan during Bolus Tracking.
|
FDA
Z-2988-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jun 26, 2026
|
Details |
|
FUJIFILM Synapse PACS software v7.4.310; radiological image processing system
Software issue may cause the ruler measurement to be inaccurate.
|
FDA
Z-3030-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 22, 2026
|
Details |
|
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
|
FDA
Z-3012-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
FRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 5700225
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
|
FDA
Z-3039-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
|
FDA
Z-3013-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
|
FDA
Z-3037-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
|
FDA
Z-3038-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 20, 2026
|
Details |
|
Merge PACS and Merge OrthoPACS, medical imaging diagnostic workstation
Software issue may cause measurements to be displayed inaccurately.
|
FDA
Z-3004-2026
|
Medical Device | High |
Sep 2, 2026
Recall: Jul 22, 2026
|
Details |
|
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 722225, 722226, 722227, 722228, 722280*, 722281*, and 722282* (*These models are distributed only in China), with Software Version Number R2.2.10 and R2.1.10* (*This software version is distributed only in China); Interventional Fluoroscopic X-Ray System
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
|
FDA
Z-2931-2026
|
Medical Device | High |
Aug 19, 2026
Recall: Jul 16, 2026
|
Details |
|
Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: T
Due to off-the-shelf display port component that has audio/video failures
|
FDA
Z-2916-2026
|
Medical Device | High |
Aug 19, 2026
Recall: Jun 4, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.