| Product Recall | Source | Type | Risk | Date | Action |
|---|---|---|---|---|---|
|
ARTIS icono ceiling. Model Number: 11328100.
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
|
FDA
Z-2648-2026
|
Medical Device | High |
Jul 8, 2026
Recall: Jun 8, 2026
|
Details |
|
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
|
FDA
Z-2579-2026
|
Medical Device | High |
Jul 1, 2026
Recall: Jun 5, 2026
|
Details |
|
Philips Avalon Fetal Monitor, FM 30 Part numberM2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
|
FDA
Z-2523-2026
|
Medical Device | High |
Jul 1, 2026
Recall: May 26, 2026
|
Details |
|
Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
|
FDA
Z-2522-2026
|
Medical Device | High |
Jul 1, 2026
Recall: May 26, 2026
|
Details |
|
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.
|
FDA
Z-2574-2026
|
Medical Device | High |
Jul 1, 2026
Recall: May 27, 2026
|
Details |
|
BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
|
FDA
Z-2514-2026
|
Medical Device | High |
Jun 24, 2026
Recall: May 21, 2026
|
Details |
|
FDA
Z-2520-2026
|
Medical Device | High |
Jun 24, 2026
Recall: May 13, 2026
|
Details | |
|
FDA
Z-2519-2026
|
Medical Device | High |
Jun 24, 2026
Recall: May 13, 2026
|
Details | |
|
GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2242-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2243-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2244-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2245-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2247-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
|
FDA
Z-2246-2026
|
Medical Device | High |
Jun 3, 2026
Recall: Apr 24, 2026
|
Details |
|
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
|
FDA
Z-2165-2026
|
Medical Device | High |
May 20, 2026
Recall: Apr 14, 2026
|
Details |
|
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
|
FDA
Z-2170-2026
|
Medical Device | High |
May 20, 2026
Recall: Mar 31, 2026
|
Details |
|
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
|
FDA
Z-2163-2026
|
Medical Device | High |
May 20, 2026
Recall: Apr 14, 2026
|
Details |
|
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used are the same for each user. The watchOS app cannot be installed independently from the iOS phone app. Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
|
FDA
Z-2164-2026
|
Medical Device | High |
May 20, 2026
Recall: Apr 14, 2026
|
Details |
|
LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
|
FDA
Z-2049-2026
|
Medical Device | High |
May 13, 2026
Recall: Mar 30, 2026
|
Details |
|
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
|
FDA
Z-2048-2026
|
Medical Device | High |
May 13, 2026
Recall: Mar 30, 2026
|
Details |
|
Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product Description: The Command Module (91496) is the core of the Spacelabs patient monitoring system, providing the processing power for all basic physiologic parameters. The Command Module inserts into the side of a Spacelabs patient monitor. A variety of configurations can be selected to suit the monitoring needs of specific patients or care units in the hospital. Component: N/A
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
|
FDA
Z-2052-2026
|
Medical Device | High |
May 13, 2026
Recall: Mar 26, 2026
|
Details |
|
AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
|
FDA
Z-1947-2026
|
Medical Device | High |
May 6, 2026
Recall: Mar 27, 2026
|
Details |
|
iLet Bionic Pancreas, REF: BB1001
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
|
FDA
Z-1996-2026
|
Medical Device | High |
May 6, 2026
Recall: Mar 12, 2026
|
Details |
|
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
|
FDA
Z-1949-2026
|
Medical Device | High |
May 6, 2026
Recall: Mar 23, 2026
|
Details |
|
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
|
FDA
Z-1922-2026
|
Medical Device | High |
Apr 29, 2026
Recall: Mar 16, 2026
|
Details |
Review FDA device recalls and early alerts for equipment, software, monitors, implants, diagnostic tools and patient-use devices.
Use the device name, company, product code, UDI, model, catalog number, serial number, lot number, software version or implant card details.
Class I recalls are generally the most serious. Read the FDA notice for patient risk, affected devices, correction/removal steps and clinical action.
If the device supports or monitors health, do not stop treatment based only on a search result. Confirm with the official notice and your clinician when needed.
Use each recall result to compare device name, company, product code, model, serial, lot number, UDI and software version. A similar name is not enough; recalls often apply only to specific versions or production ranges.
Check the notice for FDA recall class, patient risk, correction, removal and clinical action. This helps you decide whether the item should be stopped immediately or handled through a normal remedy process.
Review retailer, online marketplace, state, distribution and purchase-date details when they are available. Some recalls apply only to certain stores or channels.
Use the linked official notice for final instructions such as repair, replacement, refund, disposal, software update, stop-use or company contact.
Search the product name or brand, then compare model, serial, lot, code, date, retailer and official recall number details on the detail page.
Stop using the product if the official notice says to do so, keep the label or packaging, and follow the remedy instructions for repair, refund, replacement or disposal.
Recalls can be limited to certain production dates, factories, lots, model numbers, colors, sizes, retailers or distribution areas.
No. This page helps you search public recall data. Use the linked agency or company notice as the final source before taking action.