Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
|
FDA
Z-1301-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 11, 2025
|
Details |
|
Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
|
FDA
Z-1223-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Feb 17, 2024
|
Details |
|
PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
|
FDA
Z-1225-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 6, 2026
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1292-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1279-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with mechanical and/or electrical controls, and a monopolar electrode, intended to deliver a current from a system generator directly to tissues for cutting/coagulation/ablation during endoscopic (including laparoscopic, arthroscopic) surgery.
A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
|
FDA
Z-1222-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Feb 17, 2024
|
Details |
|
AXIOS Stent and Electrocautery-Enhanced Delivery System - 6mm x 8mm; Material Number (UPN): M00553680;
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
|
FDA
Z-1159-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 19, 2025
|
Details |
|
SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
|
FDA
Z-1239-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 19, 2025
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1280-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
LotFancy Urinary Tract Infection Urine (UTI) Test Strips; PALINOIA UTI-10 PARAMETER REAGENT STRIPS; Vaunn Medical 8-in-1 Urinalysis Reagent Strip; JNW Direct 4-in-1 UTl Urinalysis Test Strips; Urinify STRENGTH; Urinify ESSENTIAL, Urinify Urinary Tract Infection (UTI)
Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.
|
FDA
Z-1257-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Nov 22, 2025
|
Details |
|
EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705
942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
|
FDA
Z-1233-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 5, 2026
|
Details |
|
NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS
To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.
|
FDA
Z-1242-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 19, 2025
|
Details |
|
PRISMAFLEX HF1000 SET, Product Code 107140; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
|
FDA
Z-1229-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 6, 2026
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1274-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
|
FDA
Z-1231-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 6, 2026
|
Details |
|
Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
|
FDA
Z-1255-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 16, 2025
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1284-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
Vortex Surgical I2 Injection Kit, VS0500
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
|
FDA
Z-1252-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 16, 2025
|
Details |
|
Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1268-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1285-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
FDA
Z-1254-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 16, 2025
|
Details | |
|
PRISMAFLEX M150 SET, Product Code 109990; Dialyzer, High Permeability With Or Without Sealed Dialysate System
There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
|
FDA
Z-1228-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 6, 2026
|
Details |
|
*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
|
FDA
Z-1163-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Dec 19, 2025
|
Details |
|
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Devices which did not undergo thermoforming could deform and lose performance.
|
FDA
Z-1277-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 7, 2026
|
Details |
|
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
|
FDA
Z-1299-2026
|
Medical Device | High |
Feb 11, 2026
Recall: Jan 9, 2026
|
Details |
Use the Today filter when you want newly published recall alerts from the current U.S. date. If no results appear, check the last 7 or 10 days because agencies may publish updates in batches.
Recent recalls can include new notices, updated classifications, expanded lots, changed risk levels and public health alerts. Sort by newest first for the live recall feed view.
High-risk notices often involve serious injury, contamination, undeclared allergens, vehicle crash hazards, fire risks or products that should not be used or consumed.
This latest recall page combines CPSC, FDA, USDA FSIS and NHTSA data so shoppers do not need to check each agency page separately.
New alerts for toys, furniture, appliances, electronics, baby products, household goods and products with injury, choking, fire or fall hazards.
Recent FDA notices can include Class I, Class II and Class III recalls, market withdrawals, safety alerts, code information and distribution areas.
FSIS alerts cover meat, poultry and egg products, including public health alerts, undeclared allergens, pathogen risks and misbranding issues.
New vehicle, tire, equipment and child-seat campaigns include defect summaries, affected units, remedy notes and NHTSA campaign numbers.
Recent recall pages are date-driven. Use the published date and recall date to understand whether an alert is new, newly classified or recently updated.
Choose food, vehicle, medical device, baby product, electronics, household or consumer product when the latest list is too broad.
Sort by highest risk when checking for urgent problems such as contamination, serious injury hazards, undeclared allergens, fire risks or vehicle safety defects.
Use the detail page and official source link before taking action. Agency notices may include exact lot codes, model ranges, remedy steps and company contacts.
The site syncs official recall sources daily and stores updated records by source agency, dataset and official recall ID.
Some agencies publish or classify a recall after the company began the recall. Use both the published date and recall date when reviewing a notice.
Match the brand, product name, package size, UPC, lot code, best-by date, allergen or contamination issue and distribution area before taking action.
Yes. Use the risk filter or highest-risk sort to surface recalls marked critical or high risk across all official sources.