FDA product recall ยท Z-0351-2027

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695 Recall Details

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Medical Device High risk Nov 11, 2026

Confirm With The Official Notice

Use this page to screen for a possible match. Use the linked official notice before you stop use, throw away, return, repair or schedule service.

Recall overview

What This Recall Notice Says

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Before you act: a similar product name is not enough. Match the identifiers, affected dates, model or lot details, and remedy instructions in the official notice.
1. Match the item

Check device name, UDI, model, lot, serial and software version.

2. Read the risk

Review the hazard, consequence, risk level and whether the notice says to stop using, stop eating or keep the item away from children.

3. Follow the remedy

Look for refund, repair, replacement, disposal, software update, dealer service or company contact instructions.

4. Verify officially

Open the official source before making a safety decision, especially for vehicles, food, baby products and medical devices.

How To Tell If Yours Is Included

Device recalls can depend on UDI, model, catalog number, lot, serial number, software version or implant card details. Match the label before changing device use.

Recall Hazard And Risk

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

Recall Remedy

Use the official notice for remedy instructions. Product recall remedies may include repair, replacement, refund, disposal, software update, dealer service or a stop-use instruction.

Affected Product Or Item

Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695

Hazard Or Reason For Recall

There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.

How To Check If This Recall Applies To You

Match The Device Label

Compare UDI, product code, model, catalog number, lot, serial number and software version before changing device use.

Compare Codes And Date Ranges

Check UPCs, lot codes, serial numbers, model numbers, best-by dates, date codes, VIN ranges, production dates or official campaign numbers when available.

Read The Hazard Before Use

Hazards can involve injury, fire, choking, contamination, undeclared allergens, crash risk, device failure, electrical shock or incorrect labeling.

Follow The Official Remedy

Confirm final instructions in the linked official notice before returning, repairing, disposing, stopping use, stopping consumption, contacting the company or scheduling service.

Key Medical Device Recall Checks

Use these checks to decide whether the notice is relevant to the item you have, then confirm with the official source.
Identity

UDI, Model, Lot And Serial

Medical device recalls can depend on UDI, product code, software version, model, lot or serial number.

Risk

Hazard, Consequence And Risk Level

Use the hazard, consequence and risk level to understand whether the product should be stopped immediately or handled through normal recall instructions.

Action

Repair, Replacement, Refund Or Disposal

Recall remedies differ by agency and product type. The official notice is the final source for repair, refund, replacement, disposal, dealer service or stop-use steps.

Source

Official Agency Notice

This page helps you review public recall details in one place. Always use the linked agency or company notice before making a safety decision.

Related Product Recalls

More recalls from the same category.
Browse Category
Product Recall Source Type Risk Date Action
FDA Z-0343-2027
Medical Device High
Nov 11, 2026 Recall: Sep 17, 2026
Details
Nextseq 550Dx Instrument REF: 20005715, FOR IN VITRO DIAGNOSTIC USE
(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.
FDA Z-0320-2027
Medical Device High
Nov 11, 2026 Recall: May 8, 2024
Details
TruSystem 7500 surgical table, Model/Catalog Number: 4091000
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
FDA Z-0350-2027
Medical Device High
Nov 11, 2026 Recall: Sep 29, 2026
Details
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
FDA Z-0317-2027
Medical Device High
Nov 11, 2026 Recall: Sep 14, 2026
Details
Senographe Pristina 3D
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
FDA Z-0328-2027
Medical Device High
Nov 11, 2026 Recall: Aug 28, 2026
Details
Mobile column TruSystem 7500 U, Product Code 1730720
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
FDA Z-0333-2027
Medical Device High
Nov 11, 2026 Recall: Sep 14, 2026
Details

Recall Detail FAQ

Short answers for checking whether this recall notice applies to your item.

How do I know if this recall affects my product?

Compare the product name, brand, model, UPC, lot code, serial number, VIN, date code, package size, retailer and distribution details with the official notice. All key details should match.

What should I do after finding a matching recall?

Stop using, feeding, driving or consuming the product when the notice says to do so, then follow the official remedy for repair, replacement, refund, disposal, service or company contact.

Is this page the official recall notice?

No. This site helps you review public recall details. Use the linked official notice as the final source for affected products and safety instructions.

Why can a recall page change after publication?

Agencies and companies may update notices with expanded products, corrected codes, new risk classifications, remedy availability or additional distribution details.