What This Recall Notice Says
Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.